A methodology for the evaluation of guiding systems.
Application to total hip arthroplasty |
|
|
No standard methodology exists for validating computer-aided surgery;
therefore system evaluation is difficult especially when no ideal clinical goal
can be specified. Thus, we have developed an approach enabling the evaluation
of both an intrinsic system accuracy and an error
resulting from the clinical application. Such a method has been applied to
the evaluation of an image-less computer-aided guidance for total hip arthroplasty. Two parallel studies of cup orientation on SawbonesTM and patients have pointed out some
critical stages of the system when clinically applied. This will allow the
improvement of the system clinical usability. |
|
|
Publication: S. Blendea, J. Troccaz,
J.N. Ravey, P. Merloz.
Image-free cup navigation inaccuracy: a two-study approach. Journal
of Computer-Aided Surgery,
2007 May; 12(3):176-180 |
|
|